Final test values related to the nominal capacity (maximum volume)
and the indicated volume steps indicated on the instrument, obtained
when instrument and distilled water are equilibrated at ambient tem-
perature (20 °C/68 °F) and with smooth operation. According to
DIN EN ISO 8655.
Volume range Volume step A* CV* Increment Recommended
µl µl ≤ ± % ≤ % µl type of tip, µl
0.5 - 10 10 1.2 0.8 0.01 0.5 - 20
5 2.0 1.5
1 8.0 4.0
1 - 20 20 1.0 0.5 0.02 0.5 - 20
10 2.0 1.0
2 8.0 3.0
5 - 100 100 0.8 0.25 0.1 2 - 200
50 1.6 0.4
10 4.0 1.5
10 - 200 200 0.8 0.25 0.2 2 - 200
100 1.4 0.4
20 4.0 1.3
15 - 300 300 0.6 0.25 0.5 5 - 300
150 1.2 0.4
30 3.0 1.2
Accuracy tolerances for the Transferpette
®
-8/-12 electronic
* A
= Accuracy, CV
= Coefficient of Variation
Accuracy Table
!
20 °C
Ex
59
Testing instructions (SOPs) and a demo version of the EASYCAL
™
4.0 calibration software are available
for download at www.brand.de.