Checking the Volume
*) = Calculation of accuracy (A%) and variation coefficient (CV%):
A% and CV% are calculated according to the formulas for
statistical control.
x
i
= Weighing results
n = Number of weighings
Z = Correction factor
(e.g., 1.0029 µl/mg
at 20 °C, 1013 hPa)
V
0
= Nominal volume
Calculation (for nominal volume)
Accuracy* Coefficient of Variation*
Standard Deviation
V – V
0
V
0
A% = · 100
Σ (x
i
– x)
2
n – 1
s = Z ·
100 s
V
CV% =
Mean value x =
Σ x
i
n
Mean volume V = x · Z
Testing instructions (SOPs) and a demo version of the EASYCAL
TM
4.0
calibration software are available for download at www.brand.de.